About Metravi 05+ mV mA with Loop Calibrator
- Provides output and measures DC Voltage and Loop Current.
- Simulates the output of analogue transducer.
- Can supply 24V DC Loop Power.
- Can produce fast and slow auto inclined wave output for 4 20mA current.
- Auto Power Off can be set by operator
- Can test switch mode and take operation in step from zero to full scale
- Displaying can be done both in mA and %
- suitable for field use
- High Accuracy of 0.05%
- Display : 5 Digits LCD
- Output Provided :-
- DC current : 0 20mA
- Resolution : 0.001mA
- Accuracy : 0.05%
- Analogue Transducer : 0 to -20mA
- Resolution : 0.001mA
- Accuracy : 0.05%
- Loop Power : 24V
- Accuracy : 10%
- Measuring Range :-
- Voltage : 0 to 28V
- Resolution : 1mV
- Accuracy : 0.02%
- Current : -1.00 to 20mA
- Resolution : 0.001mA
- Accuracy : 0.02%
- Loop Current : 0 to 24mA
- Resolution : 0.001mA
- Accuracy : 0.02%
Reliable Electric Current CalibrationDesigned for professionals, the Metravi 05+ Loop Calibrator enables precise measurement and simulation of current and voltage in control loops. Its high-resolution capabilities facilitate accurate troubleshooting and calibration, making it suitable for industrial automation, instrumentation, and maintenance applications. The rugged build ensures durability even during extended use in challenging environments.
Versatile Output and Measurement ModesSwitch seamlessly between Simulate, Source, and Measure functions for mA and mV with this handheld calibrator. The versatile output modes cater to different diagnostic and calibration tasks, reducing the need for multiple instruments. The device supports both 2-wire loop connectivity and provides loop power, enhancing its utility in varied scenarios.
FAQ's of Metravi 05+ mV mA with Loop Calibrator:
Q: How do I use the Metravi 05+ Loop Calibrator for electric current measurement?
A: Connect the calibrator using the included test leads to your circuit or device. Select the desired mode-Source, Simulate, or Measure-using the controls. The large LCD display will show real-time readings of mA or mV values within the specified ranges. Ensure the device is powered with the 9V battery before operation.
Q: What are the benefits of using this calibrator in calibration and troubleshooting?
A: The calibrator guarantees high accuracy (0.05% of reading + 2 digits) and quick response times (<1s full scale), allowing fast and precise diagnostics. Overload and fuse protection enhance safety. Its robust design and extended battery life make it suitable for field and lab environments, helping users maintain critical instrumentation efficiently.
Q: When should I use the Source, Simulate, or Measure modes?
A: Use the Source mode to provide a reliable current or voltage signal to test devices or calibrate instruments, Simulate mode to mimic sensor outputs for configuration, and Measure mode to monitor existing signals in live loops. Each mode supports distinct tasks for process control, calibration, and troubleshooting in industrial settings.
Q: Where can the Metravi 05+ Loop Calibrator be used?
A: This calibrator is ideal for industrial plants, laboratories, automation maintenance, and field sites up to 2000 meters altitude. Its wide operating and storage temperature range and rugged ABS housing allow use in varied environments throughout India and beyond.
Q: What is the process for safe operation and protection?
A: Always ensure correct connection using the supplied test leads and follow instructions in the user manual. The device features error indication on the LCD for overload, and built-in fuse protection (0.5A/250V) safeguards both the user and equipment during operation.
Q: How long can I use the calibrator continuously before changing the battery?
A: The calibrator can operate for more than 40 hours of continuous typical use on a standard 9V battery, enabling extended testing and calibration tasks without frequent battery replacement.