About Metravi UONT200A Digital Microohms Meter
- Input: 230V, ±10%, 50Hz, Single Phase AC
- Accuracy ±1% of reading
- Display: 2 nos. 3 ½ digit LED module for Resistance and Current measurement
- Output Test Voltage: 5V for 200A & 100A current output & 10V for other current outputs
- Output Cables: 20m output test lead provided
- Protection against current lead disconnection while in use. Induction Kickback Suppression Snap acting overload protectio
Precision Measurements and User FlexibilityOffering an impressive accuracy of 0.2% rdg + 2 digits and resolution as fine as 0.1, the Metravi UONT200A delivers reliable microohm readings. Adjustable test current and user calibration provide flexibility to meet distinct testing conditions. Its dual operation mode (continuous or momentary) allows tailored use, while instant response time boosts efficiency for professionals handling single specimen measurements.
Designed for Safety and PortabilityThis microohmmeter complies with CAT II 600V safety standards and features input fuse protection, ensuring secure operation in varied electrical environments. Its lightweight build (1 kg), handheld design, and included carrying case enhance portability, making it easy to transport for onsite maintenance or repairs. Battery operation or DC adaptor options further add to its convenience, supporting both indoor and field work.
FAQ's of Metravi UONT200A Digital Microohms Meter:
Q: How does the Metravi UONT200A ensure accurate low-resistance measurements?
A: The device uses a standard 4-wire Kelvin connection to eliminate lead and contact resistance errors, ensuring precise readings from 1 to 200 with a resolution of 0.1. User calibration further enhances accuracy for critical low-resistance applications.
Q: What applications is this digital microohmmeter most suitable for?
A: It is designed for measuring very low resistance values in electrical components such as circuit breakers, switch contacts, transformers, motors, and bus bars-especially where accurate verification of continuity and minimal resistance is essential.
Q: When should I use the data hold function on this instrument?
A: The data hold function is helpful when you need to record or observe a measured value without display changes. It's particularly useful during testing in tight spaces or when readings must be retained for documentation.
Q: Where can the UONT200A be operated, considering environmental conditions?
A: The meter operates reliably between 0C and 40C and in environments with humidity up to 80% RH. Its portable design and included carry case make it suitable for both laboratory and on-site industrial settings.
Q: What is the process for replacing the power source or using an external supply?
A: You can power the device using 6 x 1.5V AA batteries or a DC 9V adaptor. This dual option allows for easy replacement in remote locations and uninterrupted operation in laboratory environments with stable external power.
Q: How does the auto power off feature benefit users?
A: The auto power off function conserves battery life by switching off the instrument after inactivity, reducing the likelihood of draining power during fieldwork and extending operational readiness.
Q: What accessories are provided with the UONT200A for measurement tasks?
A: The package includes test leads, crocodile clips, a carrying case, and an instruction manual, ensuring you have all necessary components for immediate and accurate low-resistance testing.