About Metravi 01+ mV Thermocouple & RTD Source Calibrator
- Provides analogue output for eight types of thermocouples (R, S, K, E, J, T, B, N) two types of thermo-resistance
- (Cu 50 / PT-100), DC Voltages and resistance
- C and F scales selectable
- Suitable for the field use
- High Accuracy of 0.05%
- Display : 5 Digits LCD
- Output provided :-
- Voltages range : -10mV to 110mV/-100mV to 1100mV
- Resolution : 0.01mV/0.1mV, Accuracy : 0.05%
- Resistance : 0 to 400
- Resolution : 0.1, Accuracy : 0.05%
- R type thermocouple : -40C to 1760C
- Resolution : 1C, Accuracy : 0.05%
- S type thermocouple : -20C to 1760C
- Resolution : 1C, Accuracy : 0.05%
- B type thermocouple : 400C to 1800C
- Resolution : 1C, Accuracy : 0.05%
- E type thermocouple : -200C to 1000C
- Resolution : 0.1C, Accuracy : 0.05%
- K type thermocouple : -200C to 1370C
- Resolution : 0.1C, Accuracy : 0.05%
- J type thermocouple : -200C to 1200C
- Resolution : 0.1C, Accuracy : 0.05%
- T type thermocouple : -200C to 400C
- Resolution : 0.1C, Accuracy : 0.05%
- N type thermocouple : -200C to 1300C
- Resolution : 0.1C, Accuracy : 0.05%
- PT-100 sensor : -200C to 850C
- Resolution : 0.1C, Accuracy : 0.05%
- Cu50 Sensor : -50C to 150C
- Resolution : 1C, Accuracy : 0.05%
Versatile Calibration ModesThe Metravi 01+ enables users to source, simulate, and measure both thermocouple and RTD signals. This flexibility supports a variety of calibration and troubleshooting requirements in both electrical and temperature applications, making it ideal for professional technicians and engineers.
Exceptional Accuracy and ResponseOffering accuracy of 0.02% of reading plus two digits and a fast response time of less than two seconds, the calibrator ensures reliable and precise measurement results. Its resolution of 0.01 mV and 0.1C facilitates detail-oriented work in critical calibration procedures.
Reliable Performance for Extended UsePowered by a 9V battery, the Metravi 01+ delivers up to 15 hours of continuous operation. With features like low battery indication, overload protection, and compliance with IEC61010 safety standards, it combines consistent reliability with user safety during demanding applications.
FAQ's of Metravi 01+ mV Thermocouple & RTD Source Calibrator:
Q: How do I use the Metravi 01+ calibrator for thermocouple and RTD instrument calibration?
A: To calibrate instruments, connect the calibrator in the desired mode-source, simulate, or measure-using terminal or lead connections. Select the appropriate thermocouple or RTD type, set the required output signal, and verify your device's response against the calibrator's accurate LCD readings.
Q: What types of thermocouples and RTDs are supported by the Metravi 01+ calibrator?
A: The device supports thermocouple types J, K, T, E, R, S, B, N and RTDs such as Pt100 (=0.385) and Cu50, accommodating a wide range of calibration requirements in laboratories and field environments.
Q: When should I replace the battery in my Metravi 01+ calibrator?
A: Replace the 9V battery when the low battery indication appears on the LCD display. Typically, the device provides up to 15 hours of use per battery under normal operating conditions.
Q: Where can the Metravi 01+ be used for calibration tasks?
A: It can be used onsite in field applications, laboratory environments, and workshops owing to its handheld design, robust construction, and reliable battery-powered operation.
Q: What process does the calibrator use for cold junction compensation?
A: Cold junction compensation is built-in and automatic. It accounts for environmental temperature fluctuations, ensuring that thermocouple readings remain highly accurate without manual adjustments.
Q: How does using this calibrator benefit electrical and temperature calibration activities?
A: With high accuracy, multiple output modes, broad compatibility, and safety features, the Metravi 01+ streamlines calibration workflows, enhances confidence in results, and ensures user safety during use.